TIC-LF502A
Introduction
The LF502 electromagnetic flowmeter uses Faraday’s
Law of electromagnetic induction in the same way as
conventional electromagnetic flowmeters to measure
the flow rate. Position of electrodes in the LF502 is so
designed that it can be used even in a partially-filled
pipe to measure the flow rate.
Improved functional magnetic field distribution
technique enables a high-precision flow measurement
continually from low-level to fully-filled flow
conditions. This
eliminates unnecessary piping work
such as lifting the downstream pipe section to fill the
detector pipe.
Compared with flowmeters measuring the flow rate by
means of flow level, the obstruction less LF502
flowmeter does not usually allow mud, sands and other
solid sediment stay at the bottom of the detector pipe
and
is unaffected by wave or floating solids on the fluid
surface.
The AF900 hand-held terminal (HART*
1
communicator) can be used to communicate with the
flowmeter from a remote place.
*1: HART protocol (Highway Addressable Remote
Transducer) is a communication protocol for industrial
sensors recommended by the HCF
(HART
Communication Foundation).
Detector
Signal cable
Converter
Excitation cable
Digital I/O
Power supply
4-20 mAdc
Figure 1. LF502/LF232*F Configuration
Figure 2. LF502 Electromagnetic Flowmeter
for Partially-filled Pipes
Specifications
Overall Specifications
Measurement range:
Meter size Measurement range
6” (150mm)
0 – 264 GPM(std) to 0 – 1320 GPM
(0 – 60 m
3
/h to 0 – 300 m
3
/h )
8” (200mm )
0 – 484 GPM(std) to 0 – 2420 GPM
(0 – 110 m
3
/h to 0 – 550 m
3
/h )
10” (250mm )
0 – 770 GPM(std) to 0 – 3850 GPM
(0 – 175 m
3
/h to 0 – 875 m
3
/h )
12” (300mm )
0 – 1100 GPM(std) to 0 – 5500 GPM
(0 – 250 m
3
/h to 0 – 1250 m
3
/h )
14” (350mm )
0 – 1540 GPM(std) to 0 – 7700 GPM
(0 – 350 m
3
/h to 0 – 1750 m
3
/h )
16” (400mm )
0 – 1980 GPM(std) to 0 – 9900 GPM
(0 – 450 m
3
/h to 0 – 2250 m
3
/h )
20” (500mm )
0 – 3124 GPM(std) to 0 – 15620 GPM
(0 – 710 m
3
/h to 0 – 3550 m
3
/h )
24” (600mm )
0 – 4400 GPM(std) to 0 – 22000 GPM
(0 – 1000 m
3
/h to 0 – 5000 m
3
/h )
Note: Above flow rate is almost 0 – 3.28 ft/s (std) to 0 –
16.4 ft/s (0 – 1 m/s to 0 – 5 m/s ) flow velocity.
Fluid-level range:
Meter size Specification
6” to 12” m
(150 to 300m )
1-1/4”(30mm ) to fully-filled condition.
14” to 24”
(350 to 600mm )
10% of inside tube diameter to
fully-filled condition.
Note: The fully-filled condition means a 100% of inside
tube diameter.
Accuracy: ±2% FS (when measurement range is
standard)
Note: The accuracy is measured under standard
operating
conditions at Toshiba's calibration
facility.
Electromagnetic Flowmeter for
Partially-filled Pipes
LF502/LF232*F
6” to 24”
(150 to 600 mm )
LF502
LF232*F