TH809 Ear-Type Thermometer
Thank you for purchasing the thermometer. This thermometer is
designed with an advanced infrared and ambient temperature
compensation technology for instantaneous self-diagnosis and
accurate temperature measurements.
Operating Instructions:
1. Gently squeeze the opposite ends of the thermometer to
pull off the probe cap. Do not use force to remove the cap.
2. Always use a new and undamaged probe cover.
Make sure the ear canal is clean.
3. Install Probe Cover
1) Place a new probe cover on the connection ring. (See Figure 1)
NOTEMake sure to place the Adhesive Side” of probe cover Upward.
2) Align the probe with the center of probe cover. Insert the probe into the probe cover on the connection ring.
(See Figure 2)
3) Push the connection ring until the “Click” sound. This means the probe cover has been installed successfully.
NOTEThe probe cover detector is optional. If the probe cover did not connect firmly, the “ ” icon
will flash on the LCD screen. Please check the setting of the probe cover again.
Proper installation of the probe cover ensures accurate measurements.
WarningKeep the probe covers and connection ring away from children.
4. Press “ON/MEM” button to power on. The thermometer is ready for use after the ear icon stop flashing and two short
beep sound.
5. Gently pull the ear back to straighten the ear canal and snugly position the
probe into the ear canal, aiming towards the membrane of the eardrum to obtain
an accurate reading. ( Fig.4-1)
6. Measuring the ear temperatureUse the index finger to trigger. Press the “Scan”
button until you hear a long beep. ( Fig.4-2 ).
7. Power offThis device will automatically shut down after 1 minute pending to extend battery life.
NOTE
a. It is recommended that you measure 3 times with the same ear. If the 3 measurements are different, select the
highest temperature.
b. To avoid the risk of cross contamination, please clean the probe according to “Cleaning and Storage” section after
each use.
c. Clinical repeatability: 0.20°C (1 year old), 0.14°C (1~5 years old), 0.14°C (5 years old)
Switching between Fahrenheit(°F) and Celsius(°C):
In “Power Off” mode, press and hold the “SCAN” button, then press the “ON/MEM” button for 3 seconds, icon “°C
will be switched to icon “°F”. You can also use the same process to change the LCD display from °F to °C.
Memory Function:
If the reading of the thermometer is within normal temperature range between 34
o
C to 42.2
o
C (93.2
o
F to 108.0
o
F),
when the thermometer is off, the last measurement data is saved into memory. Press the “ON/MEM” button
again to see the temperature stored.
Cleaning and Storage:
The probe is the most delicate part of the thermometer.
Use with care when cleaning the lens to avoid damage.
Storage temperature Range: It should be stored at room temperature between -20~+50°C, RH85%
Keep the unit dry and away from any liquids and direct sunlight.
The Probe should not submerge into any liquids.
** If device is accidentally used without probe cover, clean the probe as follows:
a. Please use the cotton swab with Alcohol (70% concentration) to clean the lens( on the inside of the probe).
b. Allow the probe to fully dry for at least 1 minute.
NOTEPlease check the device if damaged once it falls. If you can’t make sure of it, please send the
complete device to the nearest retailer for recalibration.
Holding the thermometer too long may cause a higher ambient temperature reading of the
probe. This could make the body temperature measurement lower than usual.
Changing the Battery:
This device is supplied with one lithium cell (CR2032 x 1).
1
Open the battery cover: Insert a pointed object into the battery cover pick hole. At the same time, use thumb to remove battery
cover.
(See Figure 1)
2
Flip the battery out with a small screw driver (See Figure 2)
3
Insert the new battery under the metal hook
on the left side and press the right side of the battery down until you hear a “click”.
(See Figure 3)
4
Replace the battery cover
The positive (+) side Up and the negative (-) side pointed Down.
Specifications:
; Temperature measurement range: 34~42.2°C (93.2~108°F)
; Operating temperature range: 10~40°C (50~104°F)
; Storage temperature Range: It should be stored at room temperature between -20~+50°C , RH85%
Transportation temperature shall be less than 70°C, RH95%
; Compliance with ASTM E1965-98, EN12470-5:2003 Clinical thermometers-Part 5:Performance of infra-red ear
thermometers(with maximum device),IEC/EN60601-1-2(EMC),IEC/EN60601-1(Safety) standards.
; Accuracy:+/-0.2
o
C (0.4
o
F) during 35.5~42°C (95.9~107.6°F), +/-0.3
o
C (0.5
o
F) for other range.
; This mode converts the ear temperature to display its “oral equivalent.”
(according to the result of the clinical evaluation)
Figure 4-1 Figure 4-2
Figure 1 Figure 2
Figure 1 Figure 2 Figure 3