HP 16500A Service Manual
Download Service manual of HP 16500A Analytical Instruments, Measuring Instruments for Free or View it Online on All-Guides.com.
Brand: HP
Category: Analytical Instruments , Measuring Instruments , Multimeter
Type: Service manual
Model: HP 16500A , HP 16501A LOGIC
Pages: 144
Performing the All System Test
To perform the All System Test, proceed as follows:
1. Install a formatted disk into each disk drive (be sure to verify that the disks are not write protected).
Install the RS-232C loopback connector on the microprocessor RS-232C port. Remove any
connections to the input BNC.
2. Access the Mainframe Test menu, as described in "Accessing the Mainframe Test System," in this
section.
3. Touch the All System Tests field. The instrument performs all the tests that are listed on the
screen except the Color Display Test and the Expander Test. See figure 6-8.
When the tests are completed, the status advisory for each test changes from UNTESTED to either
PASSED or FAILED. The All System Tests advisory changes from UNTESTED to TESTED. Note that
the Color Display Test and the Expander Test remain UNTESTED.
If any test fails, refer to section 6-7, "Troubleshooting."
Figure 6-8. All System Test
HP 16500A/16501A - Service
6-11