Using Shock Advisory Mode
4-10
4 Connect the electrodes to the electrode adapter cable. The electrodes are cor-
rectly connected when the locking ring is twisted, locking the ears of the con-
nector to the adapter cable.
Be sure to correctly connect the electrodes to the electrode adapter cable.
Incorrectly connecting the electrodes to the adapter cable can result in a
failure to deliver energy to the patient.
C. Press the Analyze Button
When you place the HP CodeMaster 100 in Advisory mode, the monitor displays
the following message:
FOR ANALYSIS, 200J
PRESS ANALYZE. SELECTED
Begin analyzing the patient’s ECG by pressing Analyze on the instrument front
panel. See Figure 4-3.