0DLQWHQDQFH
CodeMaster 100
13-14
Charging the Battery
Insert battery into the HP M2480B battery support system charging bay. The
indicator appears, to indicate that the battery is charging.
The battery is fully charged when the indicator appears. A fully
discharged HP M2476B 2.5 amp-hour battery takes approximately 2 hours to
charge (a fully discharged HP M2477B 4.0 amp-hour battery takes
approximately 3 hours to charge).
It may take longer to charge the battery pack if the temperature of the battery
is not between 10°C and 40°C. If it is outside these temperatures, the battery
support system allows the battery to warm up or cool down before charging.
A flashing indicator indicates a problem with the charging bay next
to the indicator. Notify service personnel and do not use that charging bay.
The other charging bay is not affected and you can continue to use it.
If the indicator appears, the battery did not charge. Remove the
battery from service immediately and notify service personnel.
If the indicator ( ) flashes, a battery pack capacity test is
recommended.
CHARGING
READY
FAULT
WARNING
FAULT
NOTE
START