CALCulate:AVERage:MINimum?
Read the minimum value found during a min-max operation. The
multimeter clears the value when min-max is turned on, when power
has been off, or after a remote interface reset. [Stored in volatile memory]
CALCulate:AVERage:MAXimum?
Read the maximum value found during a min-max operation. The
multimeter clears the value when min-max is turned on, when power
has been off, or after a remote interface reset. [Stored in volatile memory]
CALCulate:AVERage:AVERage?
Read the average of all readings taken since min-max was enabled.
The multimeter clears the value when min-max is turned on, when
power has been off, or after a remote interface reset. [Stored in volatile
memory]
CALCulate:AVERage:COUNt?
Read the number of readings taken since min-max was enabled. The
multimeter clears the value when min-max is turned on, when power
has been off, or after a remote interface reset. [Stored in volatile
memory]
CALCulate:NULL:OFFSet {<
value>|MINimum|MAXimum}
Store a null value in the multimeter’s Null Register. You must turn on the math
operation before writing to the math register. You can set the null value to any
number between 0 and ±
120% of the highest range, for the present
function.
MIN = –120% of the highest range. MAX = 120% of the highest
range. [Stored in volatile memory]
CALCulate:NULL:OFFSet? [MINimum|MAXimum]
Query the null value.
CALCulate:DB:REFerence {<
value
>|MINimum|MAXimum}
Store a relative value in the dB Relative Register. You must turn on the math
operation before writing to the math register. You can set the relative value to
any number between 0 dBm and ±
200 dBm.
MIN = –200.00 dBm. MAX = 200.00 dBm. [Stored in volatile memory]
CALCulate:DB:REFerence? [MINimum|MAXimum]
Query the dB relative value.