HP 34401A User's Manual
Download Operation & user’s manual of HP 34401A Multimeter for Free or View it Online on All-Guides.com.
General Information
General Information
Triggering and Memory
Reading
HOLD Sensitivity:
Samples per Trigger:
Trigger Delay:
External Trigger Delay:
External Trigger Jitter:
Memory:
0.01%, 0.1%, 1%, or 10% of reading
1 to 50,000
0 to 3600 sec ( 10
µs step size )
< 1 ms
< 500
µs
512 readings
Math Functions
Null, Min/Max/Average, dB, dBm, Limit Test (with TTL output).
dBm reference resistances: 50, 75, 93, 110, 124, 125, 135, 150,
250, 300, 500, 600, 800, 900, 1000, 1200, or 8000 ohms.
Standard Programming Languages
SCPI (
Standard Commands for Programmable Instruments
)
HP 3478A Language Emulation
Fluke 8840A, Fluke 8842A Language Emulation
Remote Interface
HP-IB (IEEE-488.1, IEEE-488.2) and RS-232C
General Specifications
Power Supply:
Power Line Frequency:
Power Consumption:
Operating Environment:
Storage Environment:
Operating Altitude:
Rack Dimensions (
HxWxD):
Weight:
Safety:
EMI:
[ 1 ]
Vibration and Shock:
Warranty:
100 V / 120 V / 220 V / 240 V
±10%.
45 Hz to 66 Hz and 360 Hz to 440 Hz.
Automatically sensed at power-on.
25 VA peak ( 10 W average )
Full accuracy for 0
°C to 55°C
Full accuracy to 80% R.H. at 40°C
-40°C to 70°C
Up to 2,000 meters
88.5 mm x 212.6 mm x 348.3 mm
3.6 kg (8 lbs)
Designed to CSA 231, UL 1244,
IEC 1010-1 (1990)
MIL-461C (data on file),
MIL-T-28800E Type III, Class 5
(data on file)
3 years standard
Accessories Included
Test Lead Kit with probes, alligator, and grabber attachments.
User’s Guide, Service Guide, test report, and power cord.
[ 1 ] Slight accuracy degradation may result when subjected
to 3 V/m radiated fields.
Chapter 8 Specifications