HP 34401A User's Manual
Download Operation & user’s manual of HP 34401A Multimeter for Free or View it Online on All-Guides.com.
Automatic Trigger Delays
If you do not specify a trigger delay, the multimeter selects an
automatic delay for you. The delay is determined by function, range,
integration time, and ac filter setting.
• DC Voltage and DC Current (for all ranges):
• Resistance (2-wire and 4-wire):
• AC Voltage and AC Current (for all ranges):
• Frequency and Period:
Integration Time
NPLC ≥ 1
NPLC < 1
Trigger Delay
1.5 ms
1.0 ms
Range
100 Ω
1 kΩ
10 kΩ
100 kΩ
1 MΩ
10 MΩ
100 MΩ
Trigger Delay
(For NPLC
≥ 1)
1.5 ms
1.5 ms
1.5 ms
1.5 ms
15 ms
100 ms
100 ms
AC Filter
Slow
Medium
Fast
Trigger Delay
7.0 sec
1.0 sec
600 ms
Remote or single/external trigger
Range
100 Ω
1 kΩ
10 kΩ
100 kΩ
1 MΩ
10 MΩ
100 MΩ
Trigger Delay
(For NPLC < 1)
1.0 ms
1.0 ms
1.0 ms
1.0 ms
10 ms
100 ms
100 ms
AC Filter
Slow
Medium
Fast
Trigger Delay
1.5 sec
200 ms
100 ms
Front panel with auto trigger ON
Trigger Delay
1.0 sec
Remote or single/external trigger
Trigger Delay
0 sec
Front panel with auto trigger ON
3
Chapter 3 Features and Functions
Triggering
81