Programming Language Selection
You can select one of three languages to program the multimeter from
the selected remote interface. The language is
SCPI when the multimeter is
shipped from the factory.
• Select one of the following: SCPI, HP 3478A, or Fluke 8840A.
• The language selection is stored in non-volatile memory, and does not
change when power has been off or after a remote interface reset.
• There are certain restrictions to be aware of when you are selecting
the interface language (see also “Remote Interface Selection,” on
page 92). The HP 3478A and Fluke 8840A/8842A languages are not
supported on the
RS-232 interface.
• Front-Panel Operation:
5: LANGUAGE (I/O MENU)
See also “To Select the Programming Language,” on page 165.
• Remote Interface Operation:
L1 select SCPI language
L2 select HP 3478A language
L3 select Fluke 8840A language
SCPI Language
HP 3478A Language
Fluke 8840A Language
HP-IB / 488
X
X
X
RS-232
X
Not allowed
Not allowed
Chapter 3 Features and Functions
Remote Interface Configuration
94