3-74
Performance Tests
PDH Frequency Measurement and Looped Clock (Options UKK, [USB],
UKJ, [USA], UKL, [USC], UKN, [USE])
Figure 3-33 Receiver Frequency Measurement Test Setup
3. Set the Synthesizer to: FREQUENCY - 1024.000 kHz sinewave AMPLITUDE
- +10dBm.
4. Set the frequency counter to RATIO A/B.
5. Press and set up the display
as shown opposite.
6. Press and set up the
display as shown opposite.
RECEIVE
TRANSMIT