3-148
Performance Tests
Transmitted Jitter/Wander Amplitude Accuracy (Option A3K [A3Q])
5. Select and setup the
Reference HP 37717C display as shown
opposite.
6. Select and setup
the Reference HP 37717C display as shown
opposite.
7. Select
and setup the
Reference HP 37717C display as shown
opposite.
8. Press and check that at the
end of the measurement period (10
seconds) the
display PEAK-PEAK result is
80.000 BITS ±5.5 BITS.
9. Disconnect all the test equipment.
RECEIVE
PDH
RECEIVE
JITTER
RESULTS
TIMING CONTROL
RUN/STOP
RESULTS
WANDER
WANDER