3-250
Performance Tests
Performance Test Record
PDH Frequency Measurement and
Looped Clock
3-75 Step 8: Frequency 2.047986 MHz 2.048014 MHz
Step 9: Offset (0 ppm) -7 ppm +7 ppm
Step 10: Frequency 2.048190 MHz 2.048219 MHz
Step 11: Offset (+100 ppm) +93 ppm +107 ppm
Step 12: Frequency 2.047781 MHz 2.047810 MHz
Step 13: Offset (-100 ppm) -93 ppm -107 ppm
Frequency Counter Pass/Fail
Step 14: 704 kb/s (0 ppm) UKK
only
-7 ppm +7 ppm
704 kb/s (-100 ppm)
UKK only
-93 ppm -107 ppm
704 kb/s (+100 ppm)
UKK only
+93 ppm +107 ppm
: Offset 8 Mb/s (0 ppm) -7 ppm +7 ppm
Offset 8 Mb/s
(-100 ppm)
-93 ppm -107 ppm
Offset 8 Mb/s
(+100 ppm)
+93 ppm +107 ppm
Offset 34 Mb/s
(0 ppm)
-7 ppm +7 ppm
Offset 34 Mb/s
(-100 ppm)
-93 ppm -107 ppm
Offset 34 Mb/s
(+100 ppm)
+93 ppm +107 ppm
Offset 140 Mb/s
(0 ppm)
-7 ppm +7 ppm
Offset 140 Mb/s
(-100 ppm)
-93 ppm -107 ppm
Table 11 Performance Test Record
Page
No.
Test Description Result
Min Actual Max