HP E1340A User's Manual
Download Operation & user’s manual of HP E1340A Portable Generator for Free or View it Online on All-Guides.com.
:POINts [SOURce:]SWEep:POINts <number> selects the number of points in a
frequency sweep.
The frequencies generated by the sweep are evenly spaced linearly between
the frequencies specified by
SOURce:FREQuency:STARt and STOP, or
CENTer and SPAN, inclusive.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
number numeric
see below |
MINimum | MAXimum
none
The legal range for
number, as well as the MINimum and MAXimum values, is
context-dependent. See the section “Sweep Considerations” at the beginning
of the SOURce:SWEep subsystem for a description of the dependencies.
Comments • Executable when initiated: Query form only
• Coupled command: Yes
• Related Commands: SOURce:FREQuency:CENTer, MODE,
SPAN, START, and STOP
• *RST Condition: SOURce:SWEep:POINts 1001
Example Setting the number of points in the sweep
SWE:POIN 100 Set 100 points in sweep
:TIME [SOURce:]SWEep:TIME <time> selects the duration of the sweep. The
duration is the time from the start of the sweep until when the last frequency
begins to be output.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
time numeric
see below
|
MINimum | MAXimum
seconds
With CALibration:STATe:AC OFF set: MINimum selects 0.187 mS *
(
points - 1); MAXimum selects 25.687 mS * (points - 1).
With CALibration:STATe:AC ON set: MINimum selects 0.250 mS *
(
points - 1); MAXimum selects 25.75 mS * (points - 1).
points is specified by the SOURce:SWEep:POINts command.
The above values bound the valid range for
time.
[SOURce:]SWEep:POINts
Chapter 7 HP E1340A Command Reference 175