HP E1340A User's Manual
Download Operation & user’s manual of HP E1340A Portable Generator for Free or View it Online on All-Guides.com.
*CLS *CLS clears the Standard Event Status Register, the Operation Status
Register, the Questionable Signal Register, and the error queue. This clears
the corresponding summary bits (3, 5, and 7) in the Status Byte Register.
*CLS does not affec t the enable masks of any of the Status Registers.
Comments • Executable when initiated: Yes
• Coupled command: No
• Related Commands: STATus:PRESet
• *RST Condition: none
*DMC *DMC <name>,<data> creates a macro with the specified name and assigns
zero, one, or a sequence of commands to the name. The sequence may be
composed of SCPI and/or Common Commands. The sequence may be sent
in IEEE-488.2 definite or indefinite block format or as a quoted string.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
name
character data 1 through 12 characters none
data
block data any valid command sequence none
Comments • Legal macro names must start with an alphabetic character and
contain only alphabetic, numeric, and underscore ("_") characters.
Alphabetic character case (upper vs. lower) is ignored.
The name is allowed to be the same as a SCPI command, but may be
not be the same as a Common Command. When the name is the
same as a SCPI command, the macro rather than the command will
be executed when the name is received if macro usage is enabled.
The SCPI command will be executed if macro usage is disabled.
• Executable when initiated: Yes
• Coupled command: No
• Related Commands: *EMC, *GMC, *LMC, *RMC
• *RST Condition: none; macro defintions are unaffected
• Power-On Condition: no macros are defined
Example Define macro to restart waveform
*DMC RESTART,#19ABOR;INIT Define macro
*CLS
Chapter 7 HP E1340A Command Reference 189