Omega DP6-MLPS1 User Manual

Download Instruction sheet of Omega DP6-MLPS1 Power Supply for Free or View it Online on All-Guides.com.

Brand: Omega

Category: Power Supply

Type: Instruction sheet for Omega DP6-MLPS1

Pages: 4

Download Omega DP6-MLPS1 Instruction sheet

Omega DP6-MLPS1 Instruction sheet - Page 1
1
Omega DP6-MLPS1 Instruction sheet - Page 2
2
Omega DP6-MLPS1 Instruction sheet - Page 3
3
Omega DP6-MLPS1 Instruction sheet - Page 4
4
INSTRUCTION
SHEET
Shop online at: omega.com e-mail: [email protected]
For latest product manuals: omegamanual.info
DP6-MMLPS1
Power Supply
M4533-0607
O
PROVIDES +12 VDC POWER FOR THE DP63000x SERIES
O
12 VDC OUTPUT @ 400 mA
O
EASILY ATTACHED TO BACK OF DP63000x
UL Recognized Component,
File # E313607
DESCRIPTION
The Model DP6-MLPS1 is a +12 VDC power supply designed to attach to
the rear of the DP63000x Series. The DP6-MLPS1 can be powered from an 85-
250 VAC source.
Caution: The maximum output current of the DP6-MLPS1 is 400 mA. Check the
specifications of the specific counter(s)/indicators(s) and sensors(s) being
used to ensure that total current requirements do not exceed 400 mA.
SAFETY SUMMARY
All safety related regulations, local codes and instructions that appear in the
manual or on equipment must be observed to ensure personal safety and to
prevent damage to either the instrument or equipment connected to it. If
equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
DIMENSIONS In inches (mm)
SPECIFICATIONS
1. POWER REQUIREMENTS: 85-250 VAC, 50/60 Hz, 14 VA.
2. POWER OUTPUT: +16 VDC max @ 4 mA
11.5 VDC min @ 400 mA
3. ENVIRONMENTAL CONDITIONS:
Operating Temperature: 0 to 60°C
Storage Temperature: -30 to 75°C
Operating and Storage Humidity: 85% max. (non-condensing) from 0°C
to 50°C.
Altitude: Up to 2000 meters
4. CERTIFICATIONS AND COMPLIANCES:
SAFETY
UL Recognized Component, File # E313607, UL 61010-1, CSA C22.2
No. 61010-1
Recognized to U.S. and Canadian requirements under the Component
Recognition Program of Underwriters Laboratories, Inc.
Output meets Class 2 power requirements per UL 1310.
IEC 61010-1, EN 61010-1: Safety requirements for electrical
equipment for measurement, control, and laboratory use, Part 1.
ELECTROMAGNETIC COMPATIBILITY
Emissions and Immunity to EN 61326: Electrical Equipment for
Measurement, Control and Laboratory use.
Notes:
1. Criterion A: Normal operation within specified limits.
2. Criterion B: Temporary loss of performance from which the unit self-
recovers.
5. CONSTRUCTION: High impact black plastic. Mounting hardware
included. Installation Category II, Pollution Degree 2.
6. CONNECTION: Two position terminal block which accepts one 14 AWG
wire (torque terminal screws to 5 inch-lbs. [0.56 N-m]).
7. WEIGHT: 2 oz (47 g)
CAUTION: Read complete instructions prior
to installation and operation of the unit.
CAUTION:
Risk of electric shock.
Class BEN 55011
Emissions:
0.5 cycle
Criterion A
3 V/rms
Criterion A
EN 61000-4-6
EN 61000-4-11
RF conducted interference
Voltage dip/interruptions
1 kV L-L,
Criterion BEN 61000-4-5Surge
1 kV signal
2 kV power
Criterion AEN 61000-4-4Fast transients (burst)
1 kV signal
2 kV L&N-E power
10 V/m
Criterion AEN 61000-4-3Electromagnetic RF fields
8 kV air discharge
4 kV contact discharge
Criterion AEN 61000-4-2Electrostatic discharge
Immunity to Industrial Locations:
LP0696X