YOKOGAWA MY10 User Manual

Download Operation & user’s manual of YOKOGAWA MY10 Circuit Tester, Test Equipment for Free or View it Online on All-Guides.com.

Brand: YOKOGAWA

Category: Circuit Tester , Test Equipment

Type: Operation & user’s manual

Model: YOKOGAWA MY10 , YOKOGAWA MY40

Pages: 2

Download YOKOGAWA MY10 Operation & user’s manual

YOKOGAWA MY10 Operation & user’s manual - Page 1
1
YOKOGAWA MY10 Operation & user’s manual - Page 2
2
IM MY40-S01E
2nd Edition:Feb.2009 (YMI)
MY10, MY40
Insulation Resistance Tester
User’s Supplements
Manual
For EMC Standards
This paper supplements the User’s Manual of MY10 and MY40 Insulation Resistance Tester.
This is to advise the additional explanations for EMC standards and the update postal address
for Yokogawa Europe.
Model
User’s Manual
Description
before the revision
Revised Content
MY10
IM MY10-E
6th Edition
EMC standard
EN55011 Class B Group 1
EN61326 Class B
Effect of radiation immunity
At the strength of radio-frequency
electromagnetic field of 3V/m
•Insulation resistance measurement
1st effective measuring range: ±10% of rdg
2nd effective measuring range: ±20% of rdg
Infinity and zero indications:
±1.4% or less of scale length
•AC Voltage measurement: ±10% of FS
EMC standards:
Compliant with
EN55011 Class B Group 1,
EN61326-1 Class B,
EN61326-2-2
Effect of radiation immunity
At the strength of radio-frequency
electromagnetic field of 3V/m
•Insulation resistance measurement
1st effective measuring range: ±10% of rdg
2nd effective measuring range: ±20% of rdg
Infinity and zero indications:
±1.4% or less of scale length
•AC Voltage measurement: ±10% of F.S.
MY40
IM MY40-E
6th Edition
EMC standard
EN55011 Class B Group 1
EN61326 Class B
Effect of radiation immunity
At the strength of radio-frequency
electromagnetic field of 3V/m
•Insulation resistance measurement
1st effective measuring range:
±(5% of rdg+12dgt)
2nd effective measuring range:
±(10% of rdg+12dgt)
•AC Voltage measurement:
±(5% of rdg+12dgt)
•Conductor resistance measurement:
±(10% of range)
EMC standards:
Compliant with
EN55011 Class B Group 1,
EN61326-1 Class B,
EN61326-2-2
Effect of radiation immunity
At the strength of radio-frequency
electromagnetic field of 3V/m
•Insulation resistance measurement
1st effective measuring range:
±(5% of rdg+12dgt)
2nd effective measuring range:
±(10% of rdg+12dgt)
•AC Voltage measurement:
±(5% of rdg+12dgt)
•Conductor resistance measurement:
±(10% of range)
YOKOGAWA EUROPE B. V. (THE NETHERLANDS)
Euroweg 2, 3225HD, Amersfoort THE NETHERLANDS